Recall Z-2620-2017 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge OrthoPACS software.

DateMay 24, 2017
StatusTerminated
ReasonA migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.
Root causeSoftware design
ActionThe recalling firm issued a letter dated 5/19/2017 via email on 5/24/2017 or via certified mail if they did not have an email address for the customer.
Quantity98 sites potentially have the affected versions
DistributionDistribution was made nationwide to medical facilities. There was no foreign/government/military distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall