Recall Z-2651-2017 ยท IZL System, X-Ray, Mobile

Carestream Health Inc

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures

DateJun 19, 2017 (initiated May 25, 2017)
StatusTerminated
ReasonCarestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.
Root causeProcess control
ActionCarestream Health sent the URGENT MEDICAL DEVICE CORRECTION Letters and Response Forms to the US consignees on May 25, 2017, via Certified Mail, Certified Receipt Requested. Letters to foreign accounts were distributed via e-mail or courier about the same time. Customers with questions were instructed to call 1-800-328-2910.
QuantityDomestic: 1449 units; Foreign: 1416 units
DistributionWorldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Switzerland, Taipei, Thailand, Turkey, and UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120062DRX-REVOLUTION MOBILE X-RAY SYSTEM
Carestream Health, Inc., Apr 13, 2012