Recall Z-2591-2017 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography

DateMay 8, 2017
StatusTerminated
ReasonWhen cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.
Root causeSoftware design (manufacturing process)
ActionMerge sent an Urgent Medical Device Recall letter dated May 5, 2017, to all affected customers. Customers were instructed to be aware of the following workaround: Do not have the simple close study red x icon available in the reading physicians toolbars. This will force them to set status on close which also presents the warning Not all images have been reviewed, do you want to set status anyway and close study or continue reviewing unviewed images? YOUR RESPONSE TO THIS NOTIFICATION IS REQUIRED By replying to this notification, you will be taking the necessary steps to ensure that the issu
Quantity47 sites have the potentially affected versions
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082144AMICAS PACS, VERSION 6.0
Amicas, Inc., Sep 5, 2008