Recall Z-2590-2017 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

Merge PACS software. The firm name on the label is Merge Healthcare, Hartland, WI.

DateMay 8, 2017
StatusTerminated
ReasonMerge PACS did not show unviewed images when the last view was skipped. There is a potential risk to health of a physician misdiagnosis because not all images are available for viewing.
Root causeSoftware design (manufacturing process)
ActionMerge Healthcare sent an Urgent Medical Device Recall letter dated May 5, 2017. The letter explains the issue and informs the customer to be aware of the workaround. The customer is also notified a software fix for the issue was released in Merge PACS 7.1.1 (or later). The customer is requested to ensure all users of the product are provided with the notification and if it has been further distributed, to notify downstream customers at once. The letter informs the customer a response is required 15 calendar days after receipt of the recall letter. The enclosed response asks, in part, i
Quantity47 sites potentially have the affected versions
DistributionUSA (nationwide) Distribution .There was no foreign/government/military distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082144AMICAS PACS, VERSION 6.0
Amicas, Inc., Sep 5, 2008