Recall Z-2703-2017 ยท BSZ Gas-Machine, Anesthesia

GE Healthcare, LLC

Carestation 620 A1, Carestation 650 A1, and Carestation 650c A1

DateMay 24, 2017 (initiated May 2, 2017)
StatusTerminated
ReasonCarestation 620, 650 and 650c has Potential for Elevated FiCO2 and Unexpected System Malfunction
Root causeDevice Design
ActionConsignees were sent on 5/2/2017 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 34082 dated May 2, 2017. The letter was addressed to Chief of Anesthesia, Director of Biomedical / Clinical Engineering & Health Care Administrator / Risk Manager. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction & Contact Information. GE Healthcare will correct affected products. GE Healthcare representative will contact the firms to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local
Quantity3,448 (334 US; 3,114 OUS)
DistributionUS Distribution including PR and to the states of : AL, AR, AZ, CA, CO, CT, FL GA, ID,IL, IN, KY, LA, , MA, MD, MI, MO, MS, MT, MN, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI, and UT. OUS: Algeria Argentina Australia Austria Belgium Bolivia Bosnia and Herzegovina Brazil B
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151570Carestation 620/650/650C
Datex-Ohmeda, Inc., Jan 21, 2016