Recall Z-2332-2017 ยท FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Arrow International Inc

Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features. The multiple-lumen catheter permits venous access to the central circulation by the way of the femoral, jugular, or subclavian veins. The ARROWg+ard Blue Plus antimicrobial catheter is intended to help provide protection against catheter-related infections. It is not i

DateMay 9, 2017
StatusTerminated
ReasonArrow International is recalling the affected product because the packaging may not be completely sealed, which may compromise sterility.
Root causePackaging
ActionArrow International sent an Urgent Medical Device Recall Notification Letter dated May 11, 2017, to affected customers. The firm's notification letter is requesting that customers immediately assess their current inventory and to discontinue and quarantine any product with the specific lot codes listed in the letter. In addition, customers were asked to complete the Recall Acknowledgement form and fax or email it back to Customer Service so they can receive a Returns Good Authorization Number for the product's return. Customers with questions were instructed to contact their local sales re
Quantity27,485 units distributed in U.S., 4,371 units distributed internationally
DistributionWorldwide Distribution - US (nationwide) and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993691ARROWG +ARD BLUE PLUS ANTIMICROBIAL MULTI-LUMEN CENTRAL VENOUS CATHETER
Arrow Intl., Inc., Mar 8, 2000