Recall Z-2225-2017 ยท DOF Lead, Atomic Absorption

Magellan Diagnostics, Inc.

Magellan Diagnostics LeadCare and LeadCare II test Kits. Product Usage: The LeadCare Blood Lead Test Kit is for in vitro diagnostic use only. The Test Kit is for the quantitative measurement of lead in fresh whole blood. This product is for professional use only. Sensors and Treatment Reagent of the Test Kit are specific for lead only. Contents of this LeadCare Blood Lead Test Kit must only b

DateJun 5, 2017 (initiated May 23, 2017)
StatusTerminated
ReasonUnderestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
Root causeDevice Design
ActionMagellan issued a Safety Notification letter from May 23-25, 2017 . The firm states following the FDA Safety Notice, published May 17, 2017, this notification is to advise your facility regarding a change to existing LeadCare product usage and labeling. Prior instructions for use included capillary and venous blood samples. However, because the LeadCare Testing Systems may underestimate blood lead levels and give inaccurate results when processing venous blood samples, the FDA recommends discontinuing the use of venous blood samples with any LeadCare Blood Lead Testing System (LeadCare, Lead
Quantity171,979 kits (total)= 153,447 (US)and 18532 (OUS)
DistributionWorldwide Distribution - US (Nationwide) Foreign: AUSTRALIA AUSTRIA BOLIVIA CANADA CHILE CHINA COLOMBIA DOMINICAN REPUB ENGLAND GERMANY INDIA UK ISRAEL ITALY JAPAN KOREA MALI MYANMAR SPAIN NEPAL NETHERLANDS NEW ZEALAND P.R. CHINA PANAMA US PERU PHILLIPINES POLAND PORTUG
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K971640LEADCARE BLOOD LEAD TESTING SYSTEM
Esa, Inc., Sep 9, 1997