Recall Z-2153-2017 ยท KWQ Appliance, Fixation, Spinal Intervertebral Body

Innovasis, Inc

Opteryx Variable Rescue Screw, Part #CP2214, UDI: *+M711CP22140/$$801716012*. Packaged in heat-sealed plastic bag. Product Usage: 4.5mm x 14mm Anterior Cervical Plate System Variable Rescue Screw. The Opteryx Cervical Plate System is intended as a temporary fixation device used for correction and stabilization of the cervical spine. Variable angle screws are included to allow the physician great

DateMay 23, 2017 (initiated May 5, 2017)
StatusTerminated
ReasonOpteryx Variable Rescue Screws are engraved with incorrect (larger) size (16 instead of 14).
Root causeError in labeling
ActionDistributors are being notified by e-mail, with attached copies of the written letter and forms to be printed, signed and returned to Innovasis on 5/5/17. Additional follow-up may be done via telephone and fax to return products.
Quantity21
DistributionUS Nationwide Distribution in the states of CA, UT, ID, TX, OH, and MN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061147OPTERYX ANTERIOR CERVICAL PLATE SYSTEM
Innovasis, Inc., Jun 30, 2006