ClearanceBrief
Recall Z-2598-2017 ยท HSB Rod, Fixation, Intramedullary And Accessories

Zimmer Biomet, Inc.

FEM IM NAIL 14MMDX30CM LEFT, FEM IM NAIL 14MMDX30CM RIGHT FEM IM NAIL 13MMDX32CM LEFT FEM IM NAIL 14MMDX32CM LEFT FEM IM NAIL 13MMDX32CM RIGHT FEM IM NAIL 14MMDX32CM RIGHT FEM IM NAIL 11MMDX34CM LEFT FEM IM NAIL 13MMDX34CM LEFT FEM IM NAIL 14MMDX34CM LEFT FEM IM NAIL 11MMDX34CM RIGHT FEM IM NAIL 13MMDX34CM RIGHT FEM IM NAIL 14MMDX34CM RIGHT FEM IM NAIL 11MMDX36CM LEFT FEM IM NAIL 12MMD

DateApr 28, 2017
StatusTerminated
ReasonZimmer Biomet is initiating a field action for sterile-packaged implants packaged in a single package configuration due to packaging design verification test failures.
Root causeUnder Investigation by firm
ActionZimmer Biomet initiated a voluntary recall for sterile-packaged implants packaged in a single package configuration due to packaging design verification test failures. Compromised packaging may result in surgical delays, possible infections, and revision surgeries due to infections. The firm distributed letters to customer via FED EX on 04/28/2017. Customers were instructed to do the following: 1. Review this notification and ensure affected personnel are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknow
Quantity24969
DistributionDomestic: Puerto Rico, AK AL AR AZ CA CO CT FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV WY Foreign: CANADA, ALGERIA ARGENTINA AUSTRALIA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142281ZIMMER M/DN INTRAMEDULLARY FIXATION SYSTEM
Zimmer, Inc., Oct 22, 2014