Recall Z-2444-2017 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Biomerieux Inc

VITEK 2 Gram Negative Susceptibility card (AST-GN87), REF 413943, 20 cards per carton.

DateMay 22, 2017 (initiated Apr 20, 2017)
StatusTerminated
ReasonThe integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).
Root causePackaging
ActionThe firm posted their recall notification on their customer website on 4/20/2017. The firm sent the recall notification letter to their customers via FedEx on 4/21/2017 for delivery 4/24/2017. A follow-up recall notice will be issued on 5/08/2017 which identified 3 additional lots impacted by the recall.
Quantity2,428 cartons
DistributionNationwide distribution in the U.S, including military and government sites. The product was also distributed to Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Ch
FDA recordOpen on fda.gov