Recall Z-2592-2017 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

Proteus 235

DateJun 19, 2017 (initiated Feb 9, 2017)
StatusTerminated
ReasonIt has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.
Root causeDevice Design
ActionIon Beam Application S.A. (IBA) will bring the defect into compliance: 1. Providing recommendations for user actions which, if followed, will prevent dark current. In order to prevent the error from occurring, the IBA customers are provided with a specific QA map that would detect dark current and instructions on how and when to use it (Annex 1). This guidance was included in IBAs User Notice which was sent to users on February 9, 2017. 2. IBA is developing a long term solution consisting of a new IC-cyclo (ionization chamber located at the cyclotron exit) and a Beam Absence Electronic U
Quantity18 systems ( 10 in the US)
DistributionWorldwide Distribution - US Distribution to the states of : MA, FL, VA, PA, IL, NJ, WA, TN, LA and TX., and to the countries of: Germany, South Korea, France, Czech Republic, Italy, Poland and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152224Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus, Proteus TK2)
Ion Beam Application S.A., Aug 9, 2016