Recall Z-2155-2017 ยท FPA Set, Administration, Intravascular

B Braun Medical Inc

CytoGuard Closed Luer Connector, Intravenous access 2 cartons of 50 units each (100 units per case)

DateApr 10, 2017
StatusTerminated
ReasonB. Braun is voluntarily recalling 13 lots of CytoGuard Closed Luer Connector due to a potential for some blisters to be punctured resulting in a compromised sterility barrier.
Root causeNonconforming Material/Component
ActionB. Braun mailed an Urgent Medical Device Recall Notice to affected customers to inform them of the issue. Customers were asked to determine their current inventory and to not destroy any of the product; however, product should be discontinued immediately and quarantined and return the acknowledgement form for product return.
Quantity417,600 units
DistributionUnited States Nationwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112636CYTOGUARD(TM) CLOSED LUER CONNECTOR
B.Braun Medical, Inc., Nov 26, 2011