Recall Z-2124-2017 ยท DCF Albumin, Antigen, Antiserum, Control

Beckman Coulter Inc.

Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine

DateApr 24, 2017
StatusTerminated
ReasonBeckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.
Root causeUnder Investigation by firm
ActionBeckman Coulter sent an Urgent Medical Device Recall letter dated April 24, 2017, to affected customers to inform them that Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1. Internal testing has confirmed that MA is recovering within the published insert range for Beckman Coulter Urine Protein Control Level 2. Customers are informed of the impact, actions to be taken, and resolution. Customers are instructed to complete and return the enclosed response form within 10
Quantity3,158 units total (1,957 units in US)
DistributionWorldwide Distribution - US (nationwide) and Internationally to Australia, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, France, Germany, Guatemala, India, Italy, Japan, Mayotte, Mexico, Netherlands, Oman, Philippines, Portugal, Qatar, Saudi Arabia, South Africa, Spain, Switzer
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082251SYNCHRON SYSTEMS MICROALBUMIN (MA) REAGENT, MODEL 475100
Beckman Coulter, Inc., Apr 8, 2009