Recall Z-1891-2017 ยท MAV Syringe, Balloon Inflation

PEROUSE MEDICAL

DOLPHIN INFLATION DEVICE CALIBER INFLATION DEVICE; Catalog Number: CL3030- 018SNA- 018SND 0185NA 0185ND; INTERNAL CODE NUMBER: 0185TR- 0185NA- 018SND 0185NA 0185ND

DateApr 5, 2017
StatusTerminated
ReasonComplaints regarding broken blister.
Root causePackaging change control
ActionConsignees were notified of the recall on April 5, 2017. They were informed of the issue and asked to take the following steps: 1. Inspect your stock to determine if you still have in your ownership devices among which the references and batch codes are listed in front page; 2. Control the potential damage of the blister according to the pictures above and the instructions mentioned on the labeling and the Instruction For Use: any damaged and/or cracked blister must be immediately discarded; 3. Use the compliant products without risk, after control. Indeed, the primary packaging will not dete
Quantity18,431
DistributionUS: AZ (shipped through Belgium to USA), MN, TN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042449DOLPHIN INFLATION DEVICE
Sedat S.A., Apr 18, 2005