Recall Z-2075-2017 ยท HAW Neurological Stereotaxic Instrument

Orthosoft, Inc. dba Zimmer CAS

Navitrack System - OS Knee Universal, Orthopedic Stereotaxic Instrument

DateMay 12, 2017 (initiated Jun 20, 2008)
StatusTerminated
ReasonZIMMER CAS voluntarily conducted a retrospective recall of the Navitrack System - OS Knee Universal, due to software anomalies which were corrected with a new revision in version 2.2.1.3.
Root causeSoftware design
ActionThis is a retrospective report of a correction taken on 20 June 2008. The two software issues were corrected in the new software version 2.2.1.3. There were 164 copies of the affected software, 5 of which were only temporary licenses. There were 152 of the 159 software copies that had a full license that were confirmed to have installed the updated instructions per the correction notice for a total percentage of 96%. All sales representatives were made aware of the software issues by written notice. Both software issues were corrected in a new revision of the software with no remaining effect.
Quantity152
DistributionDomestic: CA, FL, GA, IL, IN, KS, MD, MN, MS, NC, NJ, NY, OR, PA, RI, SD, TN, TX, WA, WI Foreign: Germany, Spain, Czech Republic, Netherlands, Switzerland, New Zealand, Morocco, Brazil, Australia, Austria, France, Italy, Singapore, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060336NAVITRACK SYSTEM-OS KNEE UNIVERSAL, MODEL PRO-05002
Orthosoft, Inc., Apr 28, 2006