Recall Z-2061-2017 ยท MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Respironics California Inc

Philips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617, 1076715, 1076716, 1076717, DU1053617, U1053617. Old Material No. 850008, 989805628251, 989805612101, 989805613391, 989805613661, 989805611761, 989805627411, 989805627431, 989805627441, 989805616411, 989805636631.

DateMay 18, 2017 (initiated Apr 24, 2017)
StatusTerminated
ReasonOver time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts data travelling through the cable during the high resistance event.
Root causeDevice Design
ActionThe firm, Philips, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 20, 2017 to its customers. The Customers letter inform them that Respironics California is recalling the Philips V60 Noninvasive Ventilators manufactured before 15 September 2015; that the recall is for all V60s with date of manufacture before 15 September 2015 and not for V60 Ventilators manufactured on or after 15 September 2015; of the recall reason, risk to health, how to identify devices subject to recall, and actions to be taken the customers. The customers were also instructed to complete and return the RECALL
Quantity30,516 units
DistributionWorldwide Distribution- US (nationwide) and countries of: Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Iceland, India, Indonesia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082660V60 VENTILATOR, MODEL V8000
Respironics California, Inc., Jan 9, 2009
K102985V60 VENTILATOR WITH PPV AND AUTO-TRAK+ SOFTWARE OPTIONS
Respironics, Inc., Apr 5, 2011