Recall Z-2098-2017 ยท MSS Folders And Injectors, Intraocular Lens (Iol)
Bausch & Lomb Surgical, Inc.
SoftPort Easy-Load Lens Delivery System Model: EZ-28V Is used for the folding and delivering of the LI61AO, LI61AOV, or LI61SE 3-piece lens into the eye. The SofPort Easy-Load Lens Delivery System consists of a syringe shaped body and tip with a plunger, drawer, and haptic puller. It is a sterile, disposable plastic device designed for single use only.
| Date | Apr 5, 2017 |
| Status | Terminated |
| Reason | Complaints concerning kinked haptics that were occurring during lens delivery. |
| Root cause | Device Design |
| Action | Bausch & Lomb sent a Medical Device Recall letter dated April 18, 2017 to all their consignee informing them of the recall. Consignees were instructed to review their inventory and quarantine any unused product and return to Stericycle, the firm conducting this voluntary product recall on behalf of Bausch & Lomb. Contact Stericycle to arrange for a return at 1-877-469-8081. For replacement devices call 1-800-338-2020. Customers with questions were instructed to call 1-877-479-8081 or BauschandLomb5410@Stericycle.com. QUARANTINE PRODUCT AND RETURN TO BAUSCH + LOMB According to our records, |
| Distribution | Nationwide Distribution to AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K970727 | MPORT FOLDABLE LENS PLACEMENT SYSTEM Chiron Vision Corp., Dec 17, 1997 |