Recall Z-2707-2017 ยท DQK Computer, Diagnostic, Programmable

Merge Healthcare, Inc.

The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.

DateMar 10, 2016
StatusTerminated
ReasonIf the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.
Root causeSoftware design
ActionThe recalling firm issued recall letters dated 3/10/2016, 5/3/2016, and 8/5/2016 via email or via certified mail if the email address is not available.
Quantity555 PDM's
DistributionDistribution US nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082421HEARTSUITE HEMODYNAMICS 7.10
Emageon, Inc., Dec 17, 2008