ClearanceBrief
Recall Z-2121-2017 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Medtest Holdings, Inc.

Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.

DateMay 20, 2017 (initiated May 15, 2014)
StatusTerminated
ReasonThe original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.
Root causeComponent design/selection
ActionCustomers were notified on 05/5/14by sending a technical bulletin (via email, fax, US Mail and posting on website) to all customers and distributors that had received lot 303104 of C7590, lot 303105 of C7591 or C7592 kits with either lot since product was approved to ship. All product shipped after 05/15/2014 included the updated package inserts. Any questions concerning the implementation of the new controls ranges may be directed to Pointe Scientific's Technical Service Department at 1-800-757-5313. Thank you for your continued support of Pointe Scientific.
Quantity132919 Vol vials
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.