Recall Z-2048-2017 ยท HAW Neurological Stereotaxic Instrument

Orthosoft, Inc. dba Zimmer CAS

NDI P7 Position Sensor, Stereotaxic Instrument

DateMay 8, 2017 (initiated Jun 9, 2009)
StatusTerminated
ReasonZimmer CAS voluntarily conducted a retrospective recall on specific P7 Position Sensors (Cameras), installed on Sesamoid and Sesamoid Plasty CAS workstations due to a series of these components determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage.
Root causeNonconforming Material/Component
ActionZimmer CAS sent an Field Recall Notice dated June 9, 2009. Zimmer CAS conducted a retrospective recall/Field Correction on given P7 Position Sensors (Cameras). These were installed on given Sesamoid and Sesamoid Plasty CAS workstations. A series of these components have been determined to be affected with a manufacturing issue whereby they may potentially stop functioning during usage. The failure mode consists in the camera stopping to track and interrupting the application with an error code 0X00000020 being displayed on the navigation panel. For any questions or concerns please contact the
Quantity34
DistributionWorldwide Distribution - US Distribution to the states of : KS, MS, WI, NJ, KS, IN, LA, FL and WA., and to the countries of : Canada, Australia, Singapore, Netherlands, UK, Austria, Germany, Switzerland and Czech Republic.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071929NAVITRACK SYSTEM- PARTIAL HIP RESURFACING UNIVERSAL
Orthosoft, Inc., Sep 19, 2007