Recall Z-2116-2017 ยท CIT Nadh Oxidation/Nad Reduction, Ast/Sgot

Medtest Holdings, Inc.

Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.

DateMay 18, 2017 (initiated Feb 15, 2012)
StatusTerminated
ReasonA Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not consistent with the 510k submission package insert.
Root causeError in labeling
ActionTechnical Bulletin Notification letter was sent 02/15/12 by email to all customers and distributors. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-757-5313.
Quantity119.168L across All lots
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K941132LIQUID AST (SGOT)REAGENT SET
Pointe Scientific, Inc.,, May 2, 1994