Recall Z-2657-2017 ยท DQK Computer, Diagnostic, Programmable
CardioTek BV
CardioTek EP-TRACER Software V1.x and V2.0 The EP-TRACER System is an electrophysiology measurement system used to acquire, filter, digitize, amplify, display, and record signals obtained during electrophysiological studies and related procedures. The system allows the user to monitor, display and record the signals. The system incorporates a stimulator intended to be used for diagnostic cardia
| Date | Apr 13, 2017 |
| Status | Completed |
| Reason | Software bug which allows parameters to be changed unintentionally during use. |
| Root cause | Software design |
| Action | The Firm, Schwarzer cardiotek, sent an "Urgent Field Safety Notice" dated 4/21/2017 to its consignees. The notice described the product, problem and actions to be taken. The consignees were instructed to do the following: " While backup pacing is active the software must not be operated. " In case of backup pacing disconnect all catheter connection pins and reconnect only the two required for stimulation to the dedicated backup pacing output ports (OUT1 or OUT2). " If the backup pacing parameters were changed by accident quickly switch backup pacing off and on again to reset the paramete |
| Quantity | 8 units US |
| Distribution | Worldwide Distribution: US (nationwide) to states of: CA, IL, OH, and MN; and countries of: Albania, Algeria, Argentina, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Cuba, Czech Rep, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, G |