Recall Z-2096-2017 ยท MLZ Vitrectomy, Instrument Cutter

Abbott Medical Optics Inc. (AMO)

AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and Irrigation Sleeve (Model Number NGP0020) Used with the AMO Whitestar Signature System. For ophthalmic use.

DateMay 16, 2017 (initiated Mar 30, 2017)
StatusTerminated
ReasonPackaging error. 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package.
Root causePackaging process control
ActionAn Urgent Field Safety notice was sent on 3/30/17 to inform customers that AMO has voluntarily initiated this Action because a possibility exists that, due to a packaging error, 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package. Use of a Vitrectomy Cutter that is a different size than expected could lead to the need to alter the surgical technique, including conjunctival dissection, incision enlargement and scleral sutures.
Quantity360 units total
DistributionWorldwide distribution to SA, TR, AZ, NL, PT, SG, AU, JP, and HK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081681VITRECTOMY CUTTER AND ACCESSORIES
Advanced Medical Optics, Inc., Oct 8, 2008