Recall Z-2156-2017 ยท DQK Computer, Diagnostic, Programmable

Siemens Medical Solutions USA, Inc

AXIOM Sensis, Computer, Diagnostic, Programmable. Intended to be used as a diagnostic and administrative tool supporting hemodynamic cardiac catheterizations and/or intracardiac electrophysiology studies.

DateApr 12, 2017
StatusTerminated
ReasonAn extremely dusty computer can cause problems at system start, or rarely, cause system fail. Perform a system check prior to performing exams. If the Sensis system fails, its functions can not be used. Clinical treatment may need to be terminated, restarted, or transferred to a functioning system.
Root causeMaterial/Component Contamination
ActionSiemens mailed a Customer Safety Advisory Notice (CSAN) to affected customers on 04/12/2017 to inform them of the issue. The Notice explained the issue and when it can occur, the possible risks involved, and the actions consignees should take to address the issue. For questions regarding this recall call 610-219-4834.
Quantity4,095
DistributionNationwide Distribution
FDA recordOpen on fda.gov