Recall Z-2668-2017 ยท HTX Rongeur

Exactech, Inc.

EXACTECH Equinoxe Narrow Modular Reverse Pilot-Tip Reamer // Catalog #: 321-45-38, Catalog #: 321-45-42 and Catalog #: 321-45-46

DateMay 25, 2017
StatusTerminated
ReasonInstructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.
Root causeError in labeling
ActionOn May 22, 2017 Exactech sent a Notification to all their Direct Accounts an Distributors telling them that a new updated technical techniques LIT # 718-01-30 Rev J is now available in their website (www.exac.com) or can be requested by contacting the Customer Service Representative at 800-392-2832, fax 352-378-2617 or via e-mail at Kaya@exac.com. This updated the technical techniques for implantation for Equinoxe Platform Shoulder System to clarify the proper technique for use of the Pilot Tip Reamers, i.e., to include caution statement that the surgeon should avoid applying a bending force
Quantity198
DistributionAL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerlan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.