ClearanceBrief
Recall Z-1908-2017 ยท KGX Tape And Bandage, Adhesive

Breg Inc

If Hip, Sterile Polar Dressing P/N 10680 Product Usage: These products are sterile dressings used as an insulation barrier between the Polar Care cold therapy pad and the skin. Sterile dressings are typically applied in a non-sterile environment after post-op and on occasion may be applied in a sterile environment. Breg has determined that a small percentage (approximately 0.22%) of these product

DateMay 2, 2017 (initiated Mar 30, 2017)
StatusTerminated
ReasonThey may not have been sealed prior to sterilization. Products with unsealed pouches will be non-sterile.
Root causeEmployee error
ActionBreg sent an Urgent Medical Device Correction Letter to customers on 4/13/17. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: Please take the following actions: 1) Determine if any potentially affected products are in your stock by comparing each product's Lot and Date to the Lot Numbers and Expiration Dates in the table above. Refer to Picture 2 for the location of these fields on the product packaging. 2) Once you have matched Lot and Date, inspect the product to identify any open (unsealed) pouches. Refer to Picture 1 below. 3) R
Quantity2,609 units total
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.