Recall Z-1925-2017 ยท DOF Lead, Atomic Absorption

Magellan Diagnostics, Inc.

Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010

DateMay 18, 2017 (initiated Nov 29, 2014)
StatusTerminated
ReasonLeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Root causeDevice Design
ActionMagellan issued a follow-on Recall notification related to the FDA Safety Notification dated May 17, 2017. This current recall follows FDAs recommendation to remove venous blood samples as an allowable sample type from all LeadCare products. This recall will supersede and replace the prior Recall notifications. Notifications issued by : UPS Letter (US), Registered Letter, E-mail, or Phone calls form May 23-May 25. Questions: Reba Daoust, Director of QA/RA 978-248-4811 or rdaoust@magellandx.com Refer to Firm's website: http://www.leadcare2.com/getmedia/2fd8a90f-5c64-4058-b6ab-5db13512f0f
Quantity49 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123563LEADCARE (R) ULTRA (TM) BLOOD LEAD TESTING SYSTEM
Magellan Diagnostics, Aug 20, 2013