Recall Z-2053-2017 ยท DSQ Ventricular (Assist) Bypass
Thoratec Corporation
HeartMate II LVAS with Pocket Controller; 106762 - HMII POCKET CONTROLLER, NA; UDI: 00813024011286 Intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; intended for use inside or outside the hospital.
| Date | May 24, 2017 (initiated Mar 30, 2017) |
| Status | Terminated |
| Reason | St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014. |
| Root cause | Device Design |
| Action | Urgent Medical Device Correction letters, dated March 29, 2017, were sent to customers. The letters provided information on what the current correction will involve to ensure patients are successfully able to exchange their pocket controller in an emergency situation. Patients will receive the updates at scheduled appointment. Questions regarding the notice, should be directed to Abbott HeartLine at 800-456-1477; or, your Abbott Sales Representative. |
| Quantity | 24,077 active units in US |
| Distribution | Worldwide Distribution -- US and Austria, Belgium, Brazil, Canada, Cayman Islands, Chile, Columbia, Croatia, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Kazakhstan, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Malay |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.