Recall Z-2052-2017 ยท DSQ Ventricular (Assist) Bypass

Thoratec Corporation

HeartMate II LVAS with Pocket Controller; 106015 - HMII LVAD SEALED KIT-PCTRL, NA UDI: 00813024011224 Intended to provide hemodynamic support in patients with end-stage, refractory left ventricular heart failure; intended for use inside or outside the hospital.

DateMay 24, 2017 (initiated Mar 30, 2017)
StatusTerminated
ReasonSt. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
Root causeDevice Design
ActionUrgent Medical Device Correction letters, dated March 29, 2017, were sent to customers. The letters provided information on what the current correction will involve to ensure patients are successfully able to exchange their pocket controller in an emergency situation. Patients will receive the updates at scheduled appointment. Questions regarding the notice, should be directed to Abbott HeartLine at 800-456-1477; or, your Abbott Sales Representative.
Quantity24,077 active units in US
DistributionWorldwide Distribution -- US and Austria, Belgium, Brazil, Canada, Cayman Islands, Chile, Columbia, Croatia, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Kazakhstan, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Malay
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.