Recall Z-2097-2017 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

Proton therapy system -Proteus 235 aka Proteus Plus. A medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

DateMay 3, 2017 (initiated Apr 10, 2017)
StatusTerminated
ReasonA PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered dose.
Root causeSoftware design
ActionTarget date for software update June 30, 2017. A Field Safety Notice was issued on 4/10/2017 describing the problem and reminding users to follow guidance in the user's manual.
Quantity2
DistributionFL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152224Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus, Proteus TK2)
Ion Beam Application S.A., Aug 9, 2016