ClearanceBrief
Recall Z-2084-2017 ยท DKJ Colorimetry, Salicylate

Beckman Coulter Inc.

Salicylate Reagent (SALY 2 x 45), Catalog No. 378194. Intended for quantitative determination of salicylate concentration in human serum or plasma.

DateApr 3, 2017
StatusTerminated
ReasonBeckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Recall letter dated 4/3/17 was sent to customers to inform them that Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole, Sulfasalazine, and Sulfapyridine. Customers are informed that with the administration of these drugs, the effects to patient analytes are: - Metamizole (Dipyrone) may generate false low Cholesterol, Enzymatic Creatinine, HDL Cholesterol, Lactate, Triglycerides and Uric Acid results. -Sulfasalazine may generate false high Salicylate results. -Sulfapyridine may generate false low Salicylate results. C
Quantity50,645 units total (43,816 units in US)
DistributionWorldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Neth
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K014173SYNCHRON SYSTEMS SALICYLATE REAGENT AND CALIBRATOR
Beckman Coulter, Inc., Feb 15, 2002