Recall Z-2083-2017 ยท CDT Lipase Hydrolysis/Glycerol Kinase Enzyme, Triglycerides

Beckman Coulter Inc.

Triglycerides Reagent (TG 2 x 300), Catalog No. 445850. Intended for quantitative determination of triglycerides concentration in human serum or plasma.

DateApr 3, 2017
StatusTerminated
ReasonBeckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Recall letter dated 4/3/17 was sent to customers to inform them that Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole, Sulfasalazine, and Sulfapyridine. Customers are informed that with the administration of these drugs, the effects to patient analytes are: - Metamizole (Dipyrone) may generate false low Cholesterol, Enzymatic Creatinine, HDL Cholesterol, Lactate, Triglycerides and Uric Acid results. -Sulfasalazine may generate false high Salicylate results. -Sulfapyridine may generate false low Salicylate results. C
Quantity124,068 units total (47,920 units in US)
DistributionWorldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Neth