Recall Z-2082-2017 ยท LBS Ldl & Vldl Precipitation, Cholesterol Via Esterase-Oxidase, Hdl

Beckman Coulter Inc.

HDL Cholesterol Reagent (HDL 2 x 200), Catalog No. A15625. Intended for quantitative determination of HDL cholesterol in the high density lipoprotein fraction of human serum or plasma.

DateApr 3, 2017
StatusTerminated
ReasonBeckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Recall letter dated 4/3/17 was sent to customers to inform them that Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole, Sulfasalazine, and Sulfapyridine. Customers are informed that with the administration of these drugs, the effects to patient analytes are: - Metamizole (Dipyrone) may generate false low Cholesterol, Enzymatic Creatinine, HDL Cholesterol, Lactate, Triglycerides and Uric Acid results. -Sulfasalazine may generate false high Salicylate results. -Sulfapyridine may generate false low Salicylate results. C
Quantity91,996 units total (51,027 units in US)
DistributionWorldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Neth