Recall Z-2122-2017 ยท MCX Catheter, Coronary, Atherectomy

Cardiovascular Systems Inc

CSI, Orbital Atherectomy System (OAS) Saline Infusion Pump, Rx only, Part Number: 7-10014, Model Number: SIP-3000.

DateApr 14, 2017
StatusTerminated
ReasonCardiovascular Systems, Inc. (CSI). has initiated a recall to remove 1,396 7-10014 Saline Infusion Pumps (SIP) which were distributed by CSI between 07 April 2015 and 04 April 2017. The 7-10014 pumps may switch to stand-by during use requiring the pump to be reset prior to continuing treatment. To date, none of the associated complaints have resulted in any patient harm.
Root causeDevice Design
ActionConsignees were sent on 4/14/2017 a CSI "Urgent Medical Device Recall" letter dated 14 APRIL 2017. The letter listed the Affected Product, Recall Description, SIP Replacement and Return & contact information. Consignees were notified the pumps will be replaced and the process is anticipated to be completed 31 August 2017. Consignees may continue to use their pumps however if a yellow light fault is observed and the troubleshooting per the IFU does not resolve the fault, discontinue use of the pump and notify CSI. Consignees were requested to complete and return the Customer Acknowledgement
Quantity1,396
DistributionUS: AK, AL, AR, AZ. , CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, AUSTRIA, GERMANY, JAPAN, SWITZERLAND.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.