ClearanceBrief
Recall Z-2609-2017 ยท DBF Ferritin, Antigen, Antiserum, Control

Siemens Healthcare Diagnostics, Inc.

Dimension Vista Ferritin (FERR) Flex reagent cartridge

DateMay 5, 2017 (initiated Mar 21, 2017)
StatusTerminated
ReasonThe concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.
Root causeUnder Investigation by firm
ActionSiemens mailed an Urgent Medical Device Correction Letter to affected customers on 03/29/2017 to inform them of the issue. The notice explained the reason for the field action, Risk to Health and actions to be taken by the customer.
Quantity29,266 units
DistributionProduct was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,It
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070552DIMENSION VISTA FERRITIN FLEX REAGENT CARTRIDGE (FERR)
Dade Behring, Inc., Apr 30, 2007