Recall Z-1874-2017 ยท CBK Ventilator, Continuous, Facility Use

Newport Medical Instruments Inc

Newport Medical HT70 and HT70- Plus Ventilators, Rx Only Product Usage: The Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the Newport HT70 family of ventilators is applicable for infant, p

DateApr 10, 2017 (initiated Mar 30, 2017)
StatusTerminated
ReasonMedtronic is issuing a voluntary field corrective action for all its Newport" HT70 and Newport" HT70 Plus ventilators because they may shutdown spontaneously during normal operation without an accompanying alarm.
Root causeSoftware in the Use Environment
ActionMedtronic sent an Urgent Field Corrective Action Notice dated April 3, 2017, to all consignees via Federal Express or certified mail. The letter informed consignees of the remote potential for ventilator reset without accompanying alarm and the actions they should take. Customers were requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice. Customers were also advised that a software service update will be available in May 2017 to address this issue. Medtronic issued a Press Release related to this voluntary field action on April 5, 2017. Consignees wit
Quantity12,966
DistributionWorldwide Distribution - US (Nationwide) Internationally to Canada, Australia, Brazil, Japan, Korea, Mexico, Bahrain, Egypt, Greece Hungary, Ireland, Israel, Jordan, Kazakhstan, Lithuania, Mauritius Namibia, Norway, Pakistan, Poland, Qatar, Romania, Russian Federation South Africa, Spain, Turkey,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090888NEWPORT HT70 FAMILY OF VENTILATORS
Newport Medical Instruments, Inc., Nov 18, 2009
K111146NEWPORT HT70 VENTILATOR
Newport Medical Instruments, Inc., Dec 1, 2011