Recall Z-1770-2017 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

Combi Dockable Neurosurgery table for the Siemens MAGNETOM Aera (1.5T) and Skyra (3T) MRI Systems

DateOct 17, 2016
StatusTerminated
ReasonDue to a potential manufacturing error, a rare possibility exists for a sudden drop in table height of the Combi Dockable Table Neurosurgery.
Root causeSoftware design (manufacturing process)
ActionSiemens mailed an Customer Safety Advisory Notice on October 17, 2016, to all affected customers to inform them of the corrective action they plan to take to correct the issue. The notice also informed users of the possible risks to patients and how the corrective action will be implemented. For further questions, please call (610) 448-6461.
Quantity3 systems distributed in the U.S.
DistributionUS Distribution to: DC and MD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101347MAGNETOM AERA; MAGNETOM SKYRA
Siemens Medical Solutions USA, Inc., Oct 1, 2010