Recall Z-2340-2017 ยท IKD Cable, Electrode

Oscor, Inc.

ATAR Reusable Extension Cable, sold under brand names: ATAR Reusable Series, ATAR 53912 Reusable Series, and XI Reusable Series. Model codes: 53912 - MDTS ATAR REUSABLE, 53912A - MDTS ATAR REUSABLE, ATAR MDTS REUSABLE, ATAR-A DARK BLUE REUSABLE, ATAR-MDT REUSABLE, ATAR-MDT2 REUSABLE, ATAR-MDT2S-A, DK BLUE 30CM REUSABLE, ATAR-MDT2S-V WHITE 30CM REUSABLE, ATAR-MDTL REUSABLE, ATAR-MDTR REUSABLE,

DateApr 10, 2017 (initiated Mar 31, 2017)
StatusTerminated
ReasonThe Reusable Oscor ATAR extension cables could separate from the connector during use.
Root causeProcess control
ActionOscor sent an Important Information letter dated March 27, 2017, to all affected customers. The letter explained that they awee aware the Reusable Oscor ATAR extension cables could separate from the connector during use when instructions for use were not followed. They pointed out to consignees that if reused, they can only be reused a maximum of 2 times and must be resterilized by Oscor EtO. Customers were notified again, when the recall expanded on 06/01/2017, and were sent notifications on 06/05/2017. Instructions to customers included to examine inventory, quarantine any affected dev
Quantity19,178 units
DistributionWorldwide Distribution -US to AL, AK, CA, CO, FL, GA, ID, IL, IN, IA , KS, KY, ME, MD, MA, MI, MN, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI, DC. Foreign distribution to following countries: Macedonia, Pakistan, Austria, Canada, Guatemala, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070926ELECTRODE/EXTENSION CABLES, MODELS ATAR
Oscor, Inc., Jun 7, 2007