ClearanceBrief
Recall Z-1899-2017 ยท NDC Calculator, Drug Dose

Roche Diabetes Care, Inc.

Accu-Chek Connect Diabetes Management App versions 2.0.0, 2.0.1 and 2.1.0 for iOS and Android Product Usage: This is a digital product available for download direct to customers from the Apple App Store and the Google Play Store. The total number of downloads is not available. The number of unique users for each affected version from the date of initial distribution through 03/23/2017 is noted in

DateMar 14, 2017
StatusTerminated
ReasonA program error (bug) in the Bolus Advisor feature, which could result in incorrect bolus advice and a potential insulin over-delivery.
Root causeSoftware design
ActionRoche Diabetes Care, Inc. discovered a software issue in the Accu-Chek Connect app versions 2.0.0, 2.0.1, and 2.1.0 that can result in incorrect bolus recommendations. This software bug may result in errors in the active insulin calculation, creating the potential for a false high insulin bolus recommendation. An insulin bolus delivered based on potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Customer notices were distributed via mail on 03/30/2017, and the firm's website was also updated with recall information on that day. Customers were in
Quantity64,324
DistributionUS Nationwide in the states of: AL, AR, AZ, CA, CO, CT, DE, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150910ACCU-CHEK Connect Diabetes Management App
Roche Diabetes Care, Inc., Jun 3, 2015