Recall Z-1824-2017 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Artis zee, Angiographic x-ray system

DateMar 30, 2017 (initiated Mar 28, 2017)
StatusTerminated
ReasonSoftware error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system functions properly until the next reactivation. Following reactivation, the system no longer initializes as intended.
Root causeNonconforming Material/Component
ActionSiemens mailed a Customer Safety Advisory Notice (CSAN) AX017/17/S to affected customers on 03/28/2017 to inform them of the corrective action that will be performed to prevent a possible hazard to persons and equipment. The notice also explained the underlying issue, the impact on system operation, potential risk to patients and actions they are taking to correct the issue.
Quantity3 units distributed in U.S.
DistributionDistributed to: VA, CA, WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073290ARTIS ZEE AND ZEEGO
Siemens Medical Solutions USA, Inc., Feb 11, 2008