Recall Z-1778-2017 ยท DQK Computer, Diagnostic, Programmable

Merge Healthcare, Inc.

Merge Hemo software. The firm name on the labeling is Merge Healthcare, Hartland, WI. Merge Hemo monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure

DateApr 11, 2017 (initiated Apr 4, 2016)
StatusTerminated
ReasonUse of the software may show an incorrect value to the user when viewing the Fractional Flow Reserve (FFR) results during recording.
Root causeSoftware design
ActionMERGE sent an Urgent Medical Device Recall letter dated April 4, 2016 to all affected customers via e-mail the week of April 4, 2016. If no email address was available, the letters were issued via certified mail. A second letter dated July 11, 2016 was issued to non-responders. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to ensure that all users of the affected product be provided with the notice. Customers were also asked to reply using the enclosed form and the return addressed envelope. Customers with question
Quantity24 sites potentially have the affected versions
DistributionNationwide Distribution including AZ, CA, CO, FL, GA, KS, LA, MI, MO, OH, OK, TN, TX, VA, WA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082421HEARTSUITE HEMODYNAMICS 7.10
Emageon, Inc., Dec 17, 2008