Recall Z-2072-2017 ยท MQP Spinal Vertebral Body Replacement Device

Eden Spine Europe SA

GIZA Vertebral Body Replacement

DateMar 17, 2017
StatusTerminated
ReasonComplaints that implants have been disassembled by surgeons because of unscrewing completely the locking screw.
Root causeLabeling design
ActionA correction letter to the users was sent on March 28, 2017 Stating the following: Eden Spine Europe has become aware of extremely rare incidents of disassembly of Giza devices, which is distributed by Stryker Spine in the USA. These incidents have occurred in limited situations when a surgeon has needed to unscrew the device in order to readjust the distraction. To mitigate this risk, a more precise description of the unlocking procedure in the Giza Surgical Technique was added to prevent surgeons from fully unscrewing the locking screw. POTENTIAL HAZARD In the unlikely event the Giza de
Quantity1,839
DistributionAK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112429GIZA(R) VERTEBRAL BODY REPLACEMENT
Eden Spine, LLC, Jan 26, 2012