Recall Z-0350-2018 ยท DQK Computer, Diagnostic, Programmable

Merge Healthcare, Inc.

Fellows "Mighty 8" Power strip (Part Number 04175-0008) and TrippLite Power strip (PS-415-HGULTRA - Part Number 13-00106-00) are two components sold as part of the Merge Hemo System.

DateApr 6, 2017
StatusTerminated
ReasonMerge is conducting a hardware recall (replacement) of Fellows "Mighty 8" Power Strip and TrippLite Power Strip due to overheat of input resistor causing the power strip to smoke.
Root causeComponent design/selection
ActionMerge began notify customers via an Urgent Medical Device Recall letter on 4/6/2017 via mail. The letter identified affected product, stated the issue, and informed customers that Merge would replace affected product. A response form is requested to be returned to Merge. Questions can be sent via email to recall@merge.com
Quantity1,150 power strips
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152864Merge Hemo
Merge Healthcare Incorporated, Apr 7, 2016