Recall Z-1888-2017 ยท JSH Culture Media, Non-Selective And Differential

Remel Inc

Blood Agar (TSA w/ 5% Sheep Blood) Plate, For In Vitro Diagnostic Use, Remel, Lenexa, KS 66215 Product Usage: Blood Agar is a solid medium recommended for use in qualitative procedures for the cultivation of a wide variety of microorganisms and visualization of hemolytic reactions.

DateApr 27, 2017 (initiated Mar 24, 2017)
StatusTerminated
ReasonConfirmed complaint of surface contamination of Listeria monocytogenes.
Root causeMaterial/Component Contamination
ActionOn 3/24/2017, the recalling firm sent a recall notice to their customers explaining the reason for the recall and instructing customers to review their stock and discard any affected product. The notice also instructed customers to advise of the recall anyone they may have further distributed the recalled lot to. Attached to the recall notice is a response form which the recalling firm instructed their consignees to fill out and return.
Quantity1,159/100 plates/packs
DistributionUS Nationwide distribution in the states of : AR, AZ, CO, IA, IL, KS, KY, LA, MI, MO, MS, MT, NC, ND, NE, NM, OH, OK, SD, and TX and Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.