Recall Z-1828-2017 ยท NFJ System, Image Management, Ophthalmic

Merge Healthcare, Inc.

Merge Eye Station and Merge Eye Care PACS.

DateApr 4, 2016
StatusTerminated
ReasonDuring an antivirus program scan of the Eye Station or Eye Care PACS, the antivirus program may detect a newly captured image to be an unwanted file and delete the image permanently.
Root causeSoftware in the Use Environment
ActionThe recalling firm issued a letter dated 3/28/2016 via email the week of 4/4/2016 or via certified mail if they did not have an email address for the customer. A second letter dated 7/11/2016 was issued to nonresponders.
Quantity1,627 sites potentially have the affected versions
DistributionDistribution was nationwide to medical facilities. Foreign distribution was made to Canada, as well as to other countries. There was also government distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110006OIS SYMPHONY IMAGE MANAGMENT SYSTEM
Ophthalmic Imaging Systems, May 18, 2011
K913929DFC-1024 & DFC-512 DIGITAL IMAGING SYSTEM
Ophthalmic Imaging Systems, Nov 27, 1991