Recall Z-1831-2017 ยท NDP Accessories, Implant, Dental, Endosseous

Instradent USA, Inc.

NEODENT CM Intraoral Scanbody, Article Number: 108.139

DateApr 7, 2017 (initiated Mar 13, 2017)
StatusTerminated
ReasonDuring internal testing, Neodent determined that the body diameter (3.8mm) of the Neodent 108.139 CM Exact Intraoral Scan body is not compatible with the platform diameter of CM Implants when the implant is intraosseoussly (subcrestally) positioned.
Root causeComponent design/selection
ActionInstradent sent an Urgent Field Safety Notice dated March 13, 2017, to all affected customers via UPS. Customers were asked to take the following actions: 1. Identify and segregate the affected units from your stock. 2. If the product is still in your Inventory, return it for credit. 3. If you have used the product successfully, there is no need to take any action 4. For all cases complete and return the enclosed Customer Confirmation Form. For further questions, please call (855) 412-8883.
Quantity170 pieces
DistributionDistributed to the states of: NJ, TX, NV, CA, PA, AZ, and IL.