Recall Z-2037-2017 ยท LOX Catheters, Transluminal Coronary Angioplasty, Percutaneous

Abbott Vascular

NC Trek RX Coronary Dilatation Catheter, Part No. 1012448-06, 1012448-08, 1012448-12, 1012448-15, 1012448-20, 1012449-06, 1012449-08, 1012449-12, 1012449-15, 1012449-20, 1012449-25, 1012450-06, 1012450-08, 1012450-12, 1012450-15, 1012450-20, 1012451-06, 1012451-08, 1012451-12, 1012451-15, 1012451-20, 1012451-25, 1012452-08, 1012452-12, 1012452-15, 1012452-20, 1012453-08, 1012453-12, 1012453-15, 10

DateMay 12, 2017 (initiated Mar 22, 2017)
StatusTerminated
ReasonProducts from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon.
Root causeDevice Design
ActionAn Urgent Field Safety Notice dated 3/22/17 was sent to customers to inform them that Abbott Vascular is recalling the NC Trek RX Coronary Dilation Catheter, NC Traveler RX Coronary Dilatation Catheter, and NC Tenku RX PCTA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon. The letter informs the customers of how does the issue occur and what action is Abbott Vascular asking the customers to take. Customers with any questions are instructed to contact local A
Quantity449,661 units total (132,040 units in US)
DistributionUS and Worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103153VOYAGER NC CORONARY DILATATION CATHETER AND NC TREK RX CORONARY DILATATION CATHETER
Abbott Vascular-Cardiac Therapies, Dec 30, 2010