Recall Z-1833-2017 ยท JMG Radioimmunoassay (Two-Site Solid Phase), Ferritin

Beckman Coulter Inc.

Access Ferritin Calibrators, Part No. 33025, Ref No. B50136C

DateApr 17, 2017 (initiated Mar 10, 2017)
StatusTerminated
ReasonBeckman Coulter is recalling the Access Ferritin Calibrators kit because it contains a calibration card with an incorrect S5 calibrator concentration.
Root causeOther
ActionBeckman Coulter sent an Urgent Medical Device Recall letter dated March 10, 2017, to inform them that Beckman Coulter has determined through customer feedback that a portion of the Access Ferritin Calibrators kit lot listed above contains a calibration card with an incorrect S5 calibrator concentration. The S5 concentration value printed on the card and encoded in the barcode is 1.500 (ng/mL) rather than the correct concentration of 1500 (ng/mL). Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact Cu
Quantity889 kits total (6 kits in US)
DistributionWorldwide Distribution - US Distribution and to the countries of : Algeria, Bangladesh, Belgium, Botswana, Croatia, Czech Republic, Denmark, Djibouti, Egypt, France, Germany, Ghana, Greece, Hungary, Ireland, Italy, Kazakhstan, Lebanon, Lithuania, Morocco, Nepal, Netherlands, Oman, Pakistan, Poland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052082FERRITIN ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL 33020
Beckman Coulter, Inc., Aug 19, 2005
K926221ACCESS FERRITIN ASSAY
Bio-Rad Laboratories, Inc., Jan 13, 1993