Recall Z-1820-2017 ยท OWB Interventional Fluoroscopic X-Ray System

Philips Electronics North America Corporation

Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029; Allura Xper FD10 OR Table 722033; Allura Xper FD10/10 OR Table 722034; Allura Xper FD20/10 OR Table 722036; Allure Xper FD20/20 722038; Allura Xper FD20/20 biplane OR Table 722039

DateMar 22, 2017
StatusTerminated
ReasonDue to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Root causeUnder Investigation by firm
ActionConsignees were notified of the recall via letter sent on 3/22/2017. The letter explains the issue, actions that should be taken by the customer, and actions to be taken by Philips.
Quantity6992 (total for all devices)
DistributionNationwide, worldwide including Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031333PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.2
Philips Medical Systems North America, Inc., May 13, 2003
K133292ALLURA XPER FD; OR TABLE
Philips Medical Systems Nederland B.V., Mar 5, 2014
K141979ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES
Philips Medical Systems Netherland BV, Aug 19, 2014