Recall Z-1818-2017 ยท DZE Implant, Endosseous, Root-Form

Keystone Dental Inc

Keystone Dental Inc. PrimaConnex Straight Implant TC, WD Internal Connection 5.0 x 11.5mm. Catalog Number(s): 15636K

DateApr 13, 2017 (initiated Mar 9, 2017)
StatusTerminated
ReasonStraight PrimaConnex Implants contained Tapered PrimaConnex Implants
Root causeLabeling mix-ups
ActionKeystone Dental initiated a market removal communication by telephone to customers on 3/9/2017. Customer Relations Fax Back Forms were completed to document these calls. In addition, a letter was sent via Federal Express/or hand delivered to all distributors and customers on 3/13/2017. Product returns requested and if implanted to implement healing and prosthetic protocol. Toll Free: 866-902-9272
Quantity59 units
DistributionCA, IL, IN, FL, MA, MN, NJ, NV, PA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051614PRIMACONNEX INTERNAL CONNECTION IMPLANT SYSTEM, PRIMACONNEX INTERNAL CONNECTION PROSTHETICS
Lifecore Biomedical, Inc., Sep 9, 2005